우리는 두 가지 유형의 고급 제품을 제작하고 특성화했습니다.Tc 픽업 루프의 선폭이 다른 평면 SQUID 경도계. 이 장치는 차폐 없이 실험실 환경에서도 FLL(자속 고정 루프) 작동으로 작동했습니다. 차폐실에서 병렬형 소자의 자기장 구배 분해능은 약 0.5pT/cmHz였다.1/2 1kHz 및 2pT/cmHz에서1/2 1Hz에서. 이 장치는 차폐실에서 심전도를 기록하는 것이 가능했습니다. 약 10 pT의 QRS-복합 피크 PP/4mm가 선명하게 관찰됩니다. 메쉬형 장치의 경우 개방형 실험실 환경에서 저주파 소음의 증가가 병렬형 장치의 경우보다 적었습니다.
The copyright of the original papers published on this site belongs to IEICE. Unauthorized use of the original or translated papers is prohibited. See IEICE Provisions on Copyright for details.
부
Mizushi MATSUDA, Tsutomu MATSUURA, Koichi KATO, Hiroshi OYAMA, Amane HAYASHI, Satoru HIRANO, Shinya KURIKI, "First Derivative High-Tc SQUID Gradiometers with Narrow Structure from Single Layer YBCO Thin Film" in IEICE TRANSACTIONS on Electronics,
vol. E85-C, no. 3, pp. 677-680, March 2002, doi: .
Abstract: We have fabricated and characterized two types of high-Tc planar SQUID gradiometers having different line width of pickup loops. The device worked in flux-locked loop (FLL) operation even in laboratory environment without any shielding. A magnetic field gradient resolution of a parallel-type device in a lightly shielded room was about 0.5 pT/cmHz1/2 at 1 kHz and 2 pT/cmHz1/2 at 1 Hz. The device was possible to record magnetocardiograms in a shielded room. QRS-complex peaks of about 10 pT PP/4mm are clearly observed. For a mesh-type device, the increase of low frequency noise in the open laboratory environment was less than that for a parallel-type.
URL: https://global.ieice.org/en_transactions/electronics/10.1587/e85-c_3_677/_p
부
@ARTICLE{e85-c_3_677,
author={Mizushi MATSUDA, Tsutomu MATSUURA, Koichi KATO, Hiroshi OYAMA, Amane HAYASHI, Satoru HIRANO, Shinya KURIKI, },
journal={IEICE TRANSACTIONS on Electronics},
title={First Derivative High-Tc SQUID Gradiometers with Narrow Structure from Single Layer YBCO Thin Film},
year={2002},
volume={E85-C},
number={3},
pages={677-680},
abstract={We have fabricated and characterized two types of high-Tc planar SQUID gradiometers having different line width of pickup loops. The device worked in flux-locked loop (FLL) operation even in laboratory environment without any shielding. A magnetic field gradient resolution of a parallel-type device in a lightly shielded room was about 0.5 pT/cmHz1/2 at 1 kHz and 2 pT/cmHz1/2 at 1 Hz. The device was possible to record magnetocardiograms in a shielded room. QRS-complex peaks of about 10 pT PP/4mm are clearly observed. For a mesh-type device, the increase of low frequency noise in the open laboratory environment was less than that for a parallel-type.},
keywords={},
doi={},
ISSN={},
month={March},}
부
TY - JOUR
TI - First Derivative High-Tc SQUID Gradiometers with Narrow Structure from Single Layer YBCO Thin Film
T2 - IEICE TRANSACTIONS on Electronics
SP - 677
EP - 680
AU - Mizushi MATSUDA
AU - Tsutomu MATSUURA
AU - Koichi KATO
AU - Hiroshi OYAMA
AU - Amane HAYASHI
AU - Satoru HIRANO
AU - Shinya KURIKI
PY - 2002
DO -
JO - IEICE TRANSACTIONS on Electronics
SN -
VL - E85-C
IS - 3
JA - IEICE TRANSACTIONS on Electronics
Y1 - March 2002
AB - We have fabricated and characterized two types of high-Tc planar SQUID gradiometers having different line width of pickup loops. The device worked in flux-locked loop (FLL) operation even in laboratory environment without any shielding. A magnetic field gradient resolution of a parallel-type device in a lightly shielded room was about 0.5 pT/cmHz1/2 at 1 kHz and 2 pT/cmHz1/2 at 1 Hz. The device was possible to record magnetocardiograms in a shielded room. QRS-complex peaks of about 10 pT PP/4mm are clearly observed. For a mesh-type device, the increase of low frequency noise in the open laboratory environment was less than that for a parallel-type.
ER -